7.1 Configuring Settings
This section describes how to configure various system functions.
>
>
2
2
3
3
4
4
5
5
6
6
1
1
7
7
8
8
1
Under
[
Start backup]
, select the operation to perform when the instrument is turned back on
(
start back up).
When set to
[ON]
, the instrument will automatically resume recording when the power comes back
on after being interrupted during measurement, for example due to a power outage.
OFF
Does not use the start state retention function.
ON
Uses the start state retention function.
If triggers are being used, the instrument will enter the trigger wait state.
If measurement is resumed while in the start retention state, data that was stored in the instrument’s internal
buffer memory before the power outage will be deleted.
2
Under
[
Backlight saver]
, select how long to wait after the last key operation until turning off
the LCD backlight.
OFF
Disables the backlight saver (leave the backlight on all the time).
30 s
,
1 min
,
2 min
,
5 min
,
10 min
Turns off the backlight when no key has been operated for the set amount of time.
You can extend the backlight’s service life by choosing a setting other than
[OFF]
. Those settings also reduce
power consumption. Press any key to turn the backlight on.
•
The instrument consumes power even when the backlight is off. It is recommended to turn off
the instrument when it is not in use.
•
If the screen is not displayed while the instrument is powered on, the backlight saver may be
enabled.
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...