19
Configuring Input Channels
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Measuring voltage
This section describes how to configure settings on the individual settings screen when measuring
voltage.
You can use
[Input]
on the settings list screen to configure the settings. (See p. 50.)
Applicable modules: U8550, U8551, U8552, U8553, U8554, LR8530, LR8531, LR8532, LR8533,
LR8534
>
>
3
3
4
4
1
1
2
2
1
Select the module (Unit) and channel to configure and select the checkbox.
Measurement will not be performed for channels whose checkboxes are not selected.
2
Select the waveform display color.
×
(OFF), 24 colors
Select
[×]
if you wish to measure the channel but not to display its waveform or numerical values on the
screen.
3
Set the input type to
[Voltage]
.
For the U8553 and LR8533, the setting cannot be changed from
[Voltage]
.
4
Under
[Range]
, select the measurement range as appropriate for the measurement target.
(For the U8550, U8551, U8552, LR8530, LR8531, or LR8532)
10 mV
,
20 mV
,
100 mV
,
200 mV
,
1 V
,
2 V
,
10 V
,
20 V
,
100 V
,
1 to 5 V
(For the U8553 or LR8533 High Speed Voltage Unit)
100 mV
,
200 mV
,
1 V
,
2 V
,
10 V
,
20 V
,
100 V
,
1 to 5 V
(For the U8554 or LR8534 Strain Unit)
1 mV
,
2 mV
,
5 mV
,
10 mV
,
20 mV
,
50 mV
,
100 mV
,
200 mV
1
Settings and Operation
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...