157
Performing Waveform Calculations
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
7
Set the type of the waveform calculation.
Four arithmetic
Performs between channels addition, subtraction, multiplication, and division.
This setting allows you to enter channels, coefficients, and constants.
(You can also set exponents as constants.)
Aggregation
Adds the measurement data and plot its sum total.
Set the channel,
[Reset start time]
, and
[Reset time]
.
Simple average
Adds and average all measurement data from the start of measurement and plots the
results.
Set the channel,
[Reset start time]
, and
[Reset time]
.
Moving average
Calculates the average using the specified number of points while moving.
Perform averaging processing using the specified number of points for each sampled
data point and plots the results.
Set the channel, and
[Number of points]
.
Integration
Adds values obtained by multiplying measurement data by the sampling period and
plot the sum total.
Set the channel,
[Reset start time]
, and
[Reset time]
.
8
(When
[Four arithmetic]
is selected)
Specify constants, target channels, and operators.
Calculation expression
(A
∗
CHa
B
∗
CHb
C
∗
CHc
D
∗
CHd)
E
A, B, C, D, E: User-specified constants
CHa, CHb, CHc, CHd: User-specified measurement channels (up to 4 channels)
: Any one operation sign of plus (+), minus (−), multiplication (
∗
), or division(/) or a blank character
If entering a blank character, you cannot enter any further terms in the parentheses
: Any one operation sign of plus (+), minus (−), multiplication (
∗
), division(/), or exponentiation (^)
or a blank character
If entering a blank character, you cannot enter a constant.
Example: To write an expression (A
∗
CH1)
2
, enter (A
∗
CH1)^2.
7
7
When you select a calculation sign,
its constants and channels will be
displayed.
When you select a calculation sign,
its constants will be displayed.
Example: [ (100
∗
U1-1) + (50
∗
U1-2) ] + [ 10 ]
You can select a calculation channel as a target channel; however, you cannot select a calculation
channel with a number that is larger than the number of the configured calculation channel.
Example: You can set
[W1]
through [
W4]
as the calculation channel for
[W5]
.
6
Numerical and W
aveform Calculations
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...