91
Analog Triggers, Pulse Triggers, Waveform Calculation Triggers (Level, Window)
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Window triggers
This section describes how to specify a range (window) using upper and lower limit values and then
activate a trigger when the waveform moves into or out of that range.
You can activate a trigger either when the waveform enters the range (window IN) or when the
waveform exits the range (window OUT).
Window IN
Window OUT
>
>
[Unit n]
(n = 1, 2, . . .)
2
2
3
3
1
1
1
1
1
Under
[Type]
for
[Start]
or
[Stop]
, set the trigger type to
[Window]
.
A window trigger will be used.
2
Choose between
[IN]
and
[OUT]
for the waveform.
IN
Activates the trigger when the waveform enters the specified range.
OUT
Activates the trigger when the waveform exits the specified range.
When
[Condition]
is set to
[AND]
, the system will determine whether the waveform is inside the specified
range.
3
Under
[Upper]
and
[Lower]
, set the upper and lower limit values, respectively.
The range defined by the upper and lower limit values will serve as the window.
When using the scaling function, post-scaling values are displayed.
2
Trigger Function
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...