89
Analog Triggers, Pulse Triggers, Waveform Calculation Triggers (Level, Window)
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Level triggers
Level triggers activate when the waveform crosses the specified level (the trigger level).
You can set the direction in which the level is crossed (the slope).
>
>
[Unit n]
(n = 1, 2, . . .)
2
2
3
3
1
1
1
1
1
Under
[Type]
for
[Start]
or
[Stop]
, set the trigger type to
[Level]
.
A level trigger will be used.
2
Select the slope.
The trigger will activate when the level is crossed in the set direction.
↑
Activates the trigger when the level is crossed the specified level from below. (Rising)
↓
Activates the trigger will activate when the level is crossed the specified level from
above. (Falling)
When
[Condition]
is set to
[AND]
, the system will determine whether the waveform has exceeded the
specified level.
3
Under
[Level]
, set the trigger level.
The trigger will activate when the set level (temperature, voltage, etc.) is crossed.
When using the scaling function, post-scaling values are displayed.
(When a strain measurement module is used)
The instrument expresses strain in terms of micro epsilon (μ
ε
). You do not need to enter the SI
prefix micro (μ).
2
Trigger Function
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...