4.1 Setting Criteria to Start and Stop Recording
86
Enable the trigger function (set to On), select recording start/stop timing, and
set trigger criteria. Triggering is affected only by those channels for which trig-
ger criteria have been set.
Make these settings on the
[Trig & Alm]
screen.
If trigger settings have been made on the
[CH]
screen, skip steps 1 and 2.
The settings in steps 1 and 2 are interlinked with those on the
[CH]
screen.
Selecting Trigger Sources and Combinations
1
Set the
[Trigger]
to
[On]
.
2
Select the
[Timing]
setting option to determine
how triggering will affect recording.
Setting options: (
∗
default setting)
(Start and stop triggers can be set for each channel in-
dependently.)
3
Select how trigger sources are logically com-
bined using the
[Trig Source]
-
[Start]
and
[Stop]
setting options.
Setting options:(
∗
default setting)
4
(When an external trigger is applied)
Set the
[Ext trig]
-
[Start]
or
[Stop]
setting to
[On]
.
Start*
Start recording according to Start Trigger
criteria.
Stop
Stop recording according to Stop Trigger
criteria.
Start&Stop
Start and stop recording according to the
Start and Stop Trigger criteria, respec-
tively.
OR*
Triggering occurs when any one of the spec-
ified trigger source criteria is met. (includes
external trigger signal)
AND
Triggering occurs only when all of the speci-
fied trigger source criteria are met. (includes
external trigger signal)
1
2
3
4
[AND]
One waveform has crossed
above 0V as the other
crosses on the upslope
[OR]
Either waveform crosses
0V on the upslope
Start
Measurement
Trigger....... Level
Level............. 0 V
Slope
..................↑
To apply a trigger when the upslope (
↑
) of the waveform crosses zero volts:
Start
Measurement
T
T
T T
When the trigger combining logic is set
to
[AND]
If trigger criteria are already met when
you press the
START/STOP
key, no trig-
gering occurs. Triggering occurs only
after all trigger sources have ceased to
meet the criteria at once, and are sub-
sequently met again.
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......