6.3 Saving Data
115
3
6
C
hapt
e
r
6 Savi
ng & Load
ing
D
a
ta
Before measuring, configure saving on the
[Setting]
screen. Waveforms and
numerical calculation results can be saved automatically. Folders are created for
saved data according to data type, and files are automatically numbered (p.109). For
waveform data, folders named with the current date (YY-MM-DD) are created auto-
matically. Before starting to measure, confirm that Auto Save is configured correctly,
and that the CF card or USB flash drive is properly installed.
Automatic Saving
Apply
1
2
*1. Otherwise, make advanced settings as needed (p.53).
Refer to
“(As needed) You can enter a file name for the waveform files in the [File
*2.The
[CSV (real time)]
and
[CSV + Calc]
settings are not available when the
record interval is 10 ms or 20 ms. Also, when opening a CSV file in a spread-
sheet program, note that the number of rows that can be loaded at one time is
limited (p. A7). When an option other than
[Separator: Comma]
has been se-
lected on the
[System]
screen, the file extension will be .TXT rather than .CSV.
To save calculation results automatically, configure calculation settings before
starting measurement (p.131).
When saving waveforms automatically, horizontal axis display setting is re-
stricted when starting measurement, but can be freely adjusted for analysis af-
terwards (although 100, 200 and 500 ms horizontal axis settings are
automatically changed to 1 s when measurement starts, and cannot be rese-
lected while measuring).
Save methods:
Off
Automatic saving is disabled.
Waveform
(realtime) *1
Waveform data in binary format is saved to CF card while
recording.
CSV(realtime)
*1,*2
Waveform data in text format is saved to CF card while re-
cording. Saved data cannot be reloaded in the instrument.
Calc(post meas.)
*1
Calculation results are saved to CF card after recording.
When
[Repeat]
is enabled (
[On]
), calculation results for
each recording are included in the recorded data file.
When
[Repeat]
is enabled (
[On]
), calculation results are
saved for each split time.
Wa Calc
*1
Waveform data in binary format is saved while recording,
and calculation results are saved when recording stops.
CSV + Calc *1,*2
Waveform data in text format is saved while recording, and
calculation results are saved when recording stops.
Select the desired save
method from the
[Auto
Save]
setting options.
After making any other necessary settings, press the
START/STOP
key.
Data is automatically saved to the CF card or USB flash drive when measurement fin-
ishes. (“Directory Structure for Saved Data” (p. 110).)
3
Displays an estimate of the remaining time avail-
able on the attached CF card/ USB flash drive.
Select the
[Setting]
screen.
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......