9.5 Synchronous Measurements with Multiple Instruments
164
Set the slave instrument.
1
Press the
SET
key to open the
[System]
screen.
2
Set the
[
External Trg In
]
setting to
[
Trigger
]
.
3
Press the
SET
key to open the
[Trig&Alm]
screen.
4
Set as follows.
Trig Func: On
Timing: Start
Trig Source Start: OR
Ext Trig Start: On
Turn the start trigger setting for all the channels
(analog, pulse) to
Off
. Failure to do so may start the
measurement without synchronization.
Press the
START/STOP
key of the slave instrument to start the measurement (trig-
ger waiting state). In this state, press the
START/STOP
key of the master instru-
ment to start the measurement. The slave instrument will start the measurement at
the same time.
1,2
3,4