Appendix 6 Maximum Recordable Time
A
12
When saving waveform files in the form of binary data to either the instrument’s internal
memory or the CF card or USB flash drive, maximum recording times are as follows.
The recording time when data obtained through 10 analog channels, four pulse
channels, and one alarm channel (no waveform calculation channels) are recorded
is shown in the table below as an example. Maximum recording time is inversely pro-
portional to the number of recording channels.
Because the header portion of waveform files is not included in these calculations,
actual times should be expected to be about 90% of the listed values.
Day values beyond 365 are not shown in the table.
When saving data in text format (CSV), the maximum recording time is reduced to 1/
10 of the value indicated in the table or less.
Appendix 6 Maximum Recordable Time
(d: days/ h: hours/ min: minutes/ s: seconds)
Interval
Recording capacity
LR8432-20
internal memory
(7MB)
9727
(256MB)*3
9728
(512MB)
9729
(1GB)
9830
(2GB)
10 ms
32 min
19 h 37 min
1 d 15 h 14 min
3 d 6 h 29 min
6 d 12 h 58 min
20 ms
1 h 4 min
1 d 15 h 14 min
3 d 6 h 29 min
6 d 12 h 58 min
13 d 1 h 57 min
50 ms
2 h 40 min
4 d 2 h 6 min
8 d 4 h 13 min
16 d 8 h 26 min
32 d 16 h 53 min
100 ms
5 h 21 min
8 d 4 h 13 min
16 d 8 h 26 min
32 d 16 h 53 min 65 d 9 h 47 min
200 ms
10 h 43 min
16 d 8 h 26 min
32 d 16 h 53 min 65 d 9 h 47 min
130 d 19 h 35
min
500 ms
1 d 2 h 49 min
40 d 21 h 7 min
81 d 18 h 14 min
163 d 12 h 29
min
327 d 59 min
1 s
2 d 5 h 39 min
81 d 18 h 14 min
163 d 12 h 29
min
327 d 59 min
−
2 s
4 d 11 h 18 min
163 d 12 h 29
min
327 d 59 min
−
−
5 s
11 d 4 h 16 min
−
−
−
−
10 s
22 d 8 h 33 min
−
−
−
−
20 s
44 d 17 h 6 min
−
−
−
−
30 s
67 d 1 h 39 min
−
−
−
−
1 min
134 d 3 h 18 min
−
−
−
−
2 min
268 d 6 h 36 min
−
−
−
−
5 min or
more
−
−
−
−
−
Maximum Recordable Time =
Recording capacity
*1
×
Recording interval (s)
No. of recording channels
*2
×
2
*1. Recording capacity: 7
×
1024
×
1024 (for internal memory)
*2. Number of recording channels = (number of analog channels) + (number of pulse channels
×
2) + (alarm channel) + (number of waveform calculation channels)
×
4
*3 Discontinued model
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......