3.3 Setting Configuration 1 – Recording Settings
52
Select whether recording should be resumed
after the specified recording length has
elapsed or after recording is stopped by a
“stop trigger” event.
Electrical noise mixed with input signals can
be removed by digital filtering. Noise suppres-
sion is most effective for longer recordings,
where it can provide higher measurement
accuracy and less scattering. Set as needed to
suit your measurement environment.
Refer to
Set the Recording Length for Repeated or One-time Recording
Repeated recording
On
Off
Select from the
[Repeat]
setting options.
Setting options:(
∗
default setting)
Off
∗
Recording stop after one recording length (time
span).
On
Recording restarts after the specified recording
length has elapsed or after recording is stopped
by a “stop trigger” event. To manually stop re-
cording, press the
START/STOP
key.
Dead time
When repeating recording is enabled (
[Repeat]
set to
[On]
)
After a recording length has elapsed, some “dead time” is
needed for internal processing before recording can re-
sume.
Because recording is not available during the dead time,
enable continuous recording (set
[Cont]
to
[On]
) if you
need to record without data loss.
Recording
Enable Digital Filtering (noise suppression) (as needed)
Select from the
[Filter]
setting options.
Setting options:
60 Hz*, 50 Hz, Off
(
∗
default setting)
fc: cut-off frequency
When the 60 Hz digital filter is selected, optimum line noise
suppression occurs when the recording length is between
two seconds and one hour.
Noise is suppressed with either 50 Hz or 60 Hz filter selec-
tions, although the measurement timing difference between
channels may be increased, depending on the recording
length. Disabling the filter (set to Off) minimizes such timing
differences (to less than 1 ms between channels).
When disabled (set to Off), broken wire detection is also
disabled (p. 59).
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......