Appendix 12 Heat Flow
A
43
Ap
pe
nd
ix
Heat flow
Measurement using LR8432
• The heat flow is identified depending on whether the heat flow gauge indicates
positive or negative.
• The double gauge function allows you to see at a glance the relationship between
temperature and heat flow.
Heat flow (in and out) cannot be found from temperature data alone. The cause of a
temperature increase can be identified by checking the heat flow.
Heat flow sensors (Z2012, Z2013, Z2014, Z2015, Z20165, Z2017) and double-sided
Thermally Conductive Tape (Z5008) can be purchased via an authorized Hioki dis-
tributor or reseller.
Heat flow
W/m
2
0
100
50
0
Heat outflow
(heat genera-
tion)
Heat inflow (heat
absorption)
Time
The temperature increases
as the measurement target
generates heat.
The temperature increases due to
the influence of the surrounding ar-
ea.
Temperature
Temperature
Heat
flow
Using Models Z2012, Z2013, Z2014, Z2015, Z2016, or
Z2017