3.4 Setting Configuration 2 – Input Channel Settings
61
3
3
C
h
apt
er 3
Set
ting
be
fo
re M
easurement
Configure these settings for each heat flow measurement channel when mea-
suring heat flow using heat flow sensors.
Heat Flow Measurement Settings
1
Select a channel (
CH 1 to CH 10
) for the
[Channel]
setting contents.
2
To enable measurement on the channel,
select On to display the check mark
[
]
.
3
Select
[Heat]
for the
[Input]
(input type) set-
ting.
4
Select the appropriate measurement ampli-
tude range for the
[Range]
setting.
Setting options:
10 mV*, 100 mV, 1 V, 10 V, 20 V, 100 V, 1-5 V
(* default setting)
5
Enter the sensitivity constant of the sen-
sor.
(When the sensitivity constant is set, the
scaling setting for the corresponding
channel is configured.)
*
The test report that came with the Heat
Flow Sensor contains the sensor’s sensi-
tivity constant.
1 2
: ON
: OFF
3 4
5
Example
For a sensitivity constant of
0.02421 mV/W m
-2
, enter
the figures shown in the
above figure. Also enter a
unit prefix, if necessary.
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......