3.5 Display Settings
67
3
3
C
h
apt
er 3
Set
ting
be
fo
re M
easurement
The vertical display range can be defined
by upper and lower limits.
Using this method, the voltage range does
not need to be selected because the opti-
mum range is selected automatically for the
display. However, upper and lower limit set-
tings are ignored when magnification and
zero position are enabled to define the dis-
play range.
1
Select
[Pos]
for the
[Type]
setting options.
2
Select from the
[Zoom]
setting options.
Setting options:
x100, x50, x20, x10, x5, x2, x1
∗
, x1/2
(
∗
default setting)
3
Set the desired zero position
[0 pos]
as a
percentage of screen height.
Setting options:
-50 to 150%
(with x1 magnification)
Zoom
Zero-position
setting range
Zoom
Zero-position
setting range
x1/2
0 to 100%
x10
-950 to 1050%
x1
-50 to 150%
x20
-1950 to 2050%
x2
-150 to 250%
x50
-4950 to 5050%
x5
-450 to 550%
x100
-9950 to 10050%
1
2
3
Change by 1
The expansion/com-
pression range de-
pends on the selected
magnification.
50%
100%
0%
50%
100%
50%
100%
0%
0%
100%
50%
0V
Zero Position: 0%
Display range
(Zoom: x1)
Measurement
range
Default Setting
Input type
Zoom
Zero
position
Input type
Zoom
Zero
position
Voltage/heat flow
x1
50
Integration (Count)
−
−
Temperature (Tc)
x20
0
Revolution (Revolve)
x1
0
Specifying the Vertical Display Range by Upper and Lower Lim-
its (expansion/compression)
10
-10
0
5
-5
0
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......