3.4 Setting Configuration 2 – Input Channel Settings
62
If the Allowable Measurement Range of Each Range is Significantly
Exceeded
"OVER" or "-OVER" is displayed for the numerical value or cursor value and the cal-
culated value or saved data is handled as a full scale value.
Refer to
“ LR8432-20 Measurement Values” (p. A13).
Refer to “Appendix 12 Heat Flow” (p. A42) for information about heat flow.
How to Select the Measurement Range
The allowable measurement range is determined
depending on the selected measurement range
and the sensitivity constant.
The allowable measurement range is expressed
in the following equation.
Allowable measurement range of selected mea-
surement range [W/m
2
] = ± (Selected measure-
ment range / Sensitivity constant of sensor [mV/
W•m
-2
])
Example: When the measurement range is set to
10 mV and the sensitivity constant is 0.012 mV/
W•m
-2
Allowable measurement range (extent) of
selected measurement range = ± (10 mV/0.012
mV/W•m
-2
)=± 833.333 W/m
2
After setting the range and the sensitivity con-
stant, configure the other settings as shown in
the right to verify the allowable measurement
range without performing any calculations.
Allowable measurement
range upper limit
Allowable measurement
range lower limit
Display span setting
Type: Pos
Zoom:
×
1/2
0 Pos: 50
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......