3.3 Setting Configuration 1 – Recording Settings
48
Configure the horizontal axis and recording settings on the
[Setting]
screen.
To save data while measuring, auto save must be enabled beforehand (p. 53).
Select the data acquisition interval to suit your measurement objectives. Note
that shorter recording intervals restrict the maximum recording time.
3.3 Setting Configuration 1 –
Recording Settings
Move to a setting item.
Apply
Open the setting options
for the item to be set.
Select from the listed
options.
1
2
Select the
[Setting]
screen.
Specify the Recording Interval
Select from the
[Interval]
setting options.
Setting options:
10ms
∗
, 20ms, 50ms, 100ms, 200ms, 500ms, 1s, 2s, 5s,
10s, 20s, 30s, 1min, 2min, 5min, 10min, 20min, 30min,
1h
(
∗
default setting)
Some waveform peaks may not be displayable with
certain interval settings.
For example, when the
[Interval]
is 1 s and the
[Time/
DIV]
(time base) is 5 s.
5s/div
1s
Peak
Broken wire detection is automatically disabled when
the interval is set to 10 ms, 20 ms or 50 ms (p. 59).
Содержание LR8432-20
Страница 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Страница 2: ......
Страница 22: ...Operating Precautions 14...
Страница 50: ...2 5 Zero Adjustment 42...
Страница 88: ...3 7 Viewing All Channel Settings in a List 80...
Страница 104: ...4 5 About Recording Operation 96...
Страница 154: ...7 4 Waveform Calculation Formula 146...
Страница 164: ...8 3 Making System Settings 156...
Страница 232: ...Appendix 12 Heat Flow A44...
Страница 236: ...Index Index 4...
Страница 237: ......
Страница 238: ......