4.1 About LCR function
46
When a measurement value is outside the guaranteed accuracy range, “
Reference Value
”
is displayed in the error display area. When this happens, the cause is likely to be one of the
following. Check the guaranteed accuracy range in "14.2 Measurement Range and Accu-
racy" (p. 342) and change the measurement conditions or you should consider the mea-
sured values as values for reference.
• Perhaps the test signal level is too low, increase the test signal level.
• If the current measurement range (during HOLD setting) is not appropriate, set again in the AUTO
range, or change the range by manual.