12.2 Timing Chart
323
12
Chapte
r 1
2 External Cont
rol
Inputting a trigger signal from EXT I/O or touching
on the screen in CONTINUOUS measurement
mode will cause the No. 1 and No. 3 parameter judgment results for the first and second items to be output
from the EXT I/O comparator result output signal lines after measurement of all panel numbers set to be exe-
cuted on the screen. (Judgment results for the third and subsequent items are not output.)
The following shows examples of the measurement timing.
(In the timing examples, the valid edge of the TRIG signal is set to falling (ON).)
12.2.4 CONTINUOUS Measurement Mode
CONTINUOUS measurement Screen
Example: Continuous measurement using panel numbers 1, 2, and 4
*1: No.x_PARAy-HI, No.x_PARAy-IN, No.x_PARAy-LO, AND
*2: Reset when signal changes to EOM (HIGH): HIGH
Do not reset when signal changes to EOM (HIGH): Last judgment result remains
(Start-of-measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
(Comparator result output)
Last judgment result
Judgment result
(Error output)
Signal Line
Description
INDEX, EOM
For both INDEX and EOM, a transition to HIGH is performed when the first panel measurement starts
after the trigger signal is input, and a transition to LOW is performed after measurement of the last panel
is finished and the judgment result has been output. (The HIGH level is maintained during continuous
measurement.)
AND
When the judgment results of all panels are IN, LOW is output.
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......