6.3 Judging with Upper and Lower Limit Values (Comparator Measurement Mode)
205
6
Chapte
r 6
TRANSFORMER Functio
n
3
Setting Upper and Lower Limit Values as (
%) Values Relative to the Off-
set from the Reference Value (Deviation Percentage Mode)
Procedure
1
Press .
Press
to select deviation percentage
mode.
2
• In the deviation percentage mode, the calculated
value is displayed as a deviation (
%) from the refer-
ence value.
• The reference value and upperand lower limit values
are set in the same manner as in the percentage mode.
See
• The settings of the reference value and the upper
and lower limit values are common to both the per-
centage mode and deviation percentage mode.
• The
% value is calculated using the following equation:
% =
|reference value|
Calculated value - reference value
×100
TRANSFORMER Measurement Screen
Comparator Condition Setting
Changing the unit (
a/ f/ p/ n/
/ m/ None/ k/ M/ G
)
Step the units up.
Step the units down.
Press
and use the numeric keypad to
enter the reference value.
3
Settable range: -9.99999G to 9.99999G
Reference Value Setting
4
Press
to confirm the reference value.
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......