8.3 Compensating Values to Match Reference Values (Load Compensation)
237
8
Chapter 8 Error Compens
ation
Set the compensation frequency.
9
When the load compensation is valid for the set
measurement conditions, ON appears on the
LOAD parameter in the measurement Screen.
When the same compensation frequency has been set
to multiple load compensation groups, only the group
with the smallest number will be valid.
When the current measurement frequency does not
agree with the load compensation frequency, the load
compensation will be invalid and ON will not appear.
LCR Measurement Screen
2.
Use the numeric keypad to enter the compensation fre-
quency.
Settable range: DC, 1 mHz to 200 kHz
3.
Press a instrument key to confirm the setting.
4.
Press
to close the setting screen.
Compensation Frequency Setting
• When performing load compensation for DC resis-
tance measurement: Press
.
• If you make a mistake during input:
press
to cancel the input and start again.
• When you want to cancel input:
Press
to close the compensation fre-
quency setting screen.
Load Compensation Setting
1.
Press
.
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......