4.3 Setting DC Resistance Measurement
88
There are the following three methods for setting the measurement range.
4.3.5 Setting the Measurement Range
1
Setting the method for determining the measurement range
(HOLD, AUTO, JUDGE SYNC)
Selecting the HOLD or AUTO setting while the judgment synchronization setting is enabled
automatically disables the judgment synchronization setting.
Setting AUTO Ranging
AUTO
The most suitable test range is set automatically.
(This allows the most suitable measurement range to be set when, for example,
measuring a test sample whose nature is unknown.)
HOLD
The measurement range is fixed. The range is set manually.
(When the range is fixed, high-speed measurement is possible.)
JUDGE SYNC
The optimal range is set automatically based on the comparator and
BIN measurement judgment standards.
(This allows the optimal range to be fixed relative to the comparator and BIN
measurement judgment results when testing a sample whose impedance var-
ies greatly with frequency.)
Procedure
LCR Measurement Screen
1
Rdc Basic Settings