4.2 Setting Basic Settings of Measurement Conditions
55
4
Chapte
r 4
LCR Function
The constant current operation range differs depending on the test sample to be measured.
The impedance
Zm
' observed from the generator is as follows:
Accordingly, the voltage
Vm
across both leads of the sample is as follows:
Because the generator output voltage range is 5[mV] to 5[V], the CV operation range per
the above expression is
Vm
= 0.8[mV] to 0.78[V].
In low Z high accuracy mode, the output resistance
Ro
becomes 25 [
].
Ro
: Output resistance (100 [
])
Zm'
=
Ro
+
Zm
=
100
[
] -
j15.9
[
]
Vm
Zm
Vo
Zm
----------------------
15.9
Vo
101.3
--------------------------
=
=
Vo
: generator output
|
Zm'
|
15.9
[
]
×
Vo
Constant current (CC) mode setting
Measurement mode (p. 72)
Normal mode
Low Z high accuracy mode
Constant current setting range
0.01 mA to 50.00 mA
0.01 mA to 100.00 mA
Constant current accuracy
±10%rdg. ±10
A
±10%rdg. ±10
A
Output impedance
100
±10
25
±5
Testing some types of sample is not possible using constant current. In this case, the following
symbol appears on the display:
Constant current measurement will not be performed.
Change the constant current level so that it is less than or equal to the displayed Iac monitor
values.
Example: When a 1 mH impedance is measured at 1 kHz, the CC operation range can be
obtained as follows.
Sample impedance
Zm
becomes as follows:
The impedance
Zm'
observed from the generator is as follows:
Accordingly, the current
Im
across both leads of the sample is as follows:
Because the generator output voltage range is 5 [mV] to 5 [V] (see the table of page 54),
the CC operation range per the above expression is
Im
= 49.9[
A] to 49.9[mA].
In low Z high accuracy mode, the output resistance
Ro
becomes 25 [
].
Xm
2
fL
=
Zm
=
Rm
+
jXm
=
0
[
] -
j6.28
[
]
Ro
: output resistance (100 [
])
Zm'
=
Ro
+
Zm
=
100
[
] -
j6.28
[
]
Im
Vo
Zm
------------
Vo
118.1
------------------
=
=
Vo
: generator output
|
Zm'
|
Vo
100.2
[
]
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......