8.2 Short Circuit Compensation
228
Acquire the compensation values at the set measurement frequencies. Measurement frequencies can be set
for up to five points.
8.2.2 Spot Compensation
Procedure
1
Select
in the short circuit compen-
sation screen, and press
to confirm
the selection.
Short Compensation Setting
2
Short Compensation (Spot Compensation)
Use
or
to select the compensation
point you want to set or edit, and then press
.
When you do not want to perform compen-
sation:
Press .
Compensation is not performed and the compensa-
tion screen is redisplayed.
3
Spot Compensation Value Setting
Until one of these keys is pressed for input of a nu-
merical value, the previous frequency for which SPOT
compensation was performed is displayed.
• Settable range: DC, 1 mHz to 200 kHz
• If a value in excess of 200 kHz is entered, the
measurement frequency will automatically set to
200 kHz.
• If a frequency of less than 1 mHz is set, the value will
be automatically changed to 1 mHz. However, very
small values will cause the DC setting to be used.
• If you make a mistake during input:
press
to cancel the input and start again.
Use the numeric keypad to enter a frequency
for compensation.
Press
to confirm the frequency for compensation.
4
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......