This function is for outputting a measurement terminal connector error when the measurement result is high
relative to the set judgment reference value. The setting value can be set as an absolute value, and the error
is output via the measurement screen and EXT I/O. On the Measurement screen, this error is output as [Hi Z].
See
"Chapter 12 External Control" (p. 309)
The judgment reference is calculated from the nominal value (range name) of the current measurement range
and the judgment reference value as shown below.
Judgment reference = Nominal value of current measurement range × Judgment reference value (%)
5.3.3
Detecting OPEN during 2-terminal Measurement
(HIGH-Z Reject Function)
Example Current measurement range: 10 k
Judgment reference value: 150%
Judgment reference = 10 k × 1.50 = 15 k
Procedure
ANALYZER Application Settings
1
ANALYZER Measurement Screen
Press
.
2
ANALYZER Application Settings