4.4 Judging Measurement Results
106
The set reference value and upper and lower limit values are common to percentage mode
and percentage deviation mode.
5
• Settable range: -999.999% to 999.999%
Press
and use the numeric keypad to
set the upper limit value.
Set the upper limit value as a percentage relative
to the reference value.
• The actual internal operation consists of calculating the upper-limit value of comparison using the equation
given below, and comparing it to the measurement value to enable a decision to be made.
Upper limit comparison value = reference value + |reference value| ×
Upper Limit Value Setting
When you do not want to set the upper
limit:
Press
.
100
Percentage set value
6
Press
to confirm the upper limit value.
• The actual internal operation calculates the lower limit comparison value with the following equation, and when
a value that is lower than the reference value is set, the minus (-) sign is required for the percentage setting
value.
Lower limit comparison value = reference value + |reference value| ×
100
Percentage set value
• Settable range: -999.999% to 999.999%
7
Return to
Step
2
, press
, use the numeric keypad to enter the lower limit value, and
press .
8
Press
to close the setting screen.
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......