8.1 Setting Open Circuit Compensation
218
In "ALL" compensation, compensation is performed for the entire frequency range. By setting the minimum
and maximum compensation frequencies with this function, you can reduce the time required to perform the
compensation process. The DC on/off setting as well as the compensation minimum and maximum frequency
settings apply to both open and short compensation.
Compensation range limitation function
Procedure
1
Open Compensation (All Compensation)
Press
.
2
Compensation Range Limitation Setting
Turn DC open compensation on or off.
Disables DC open compensation.
Enables DC open compensation.
When you want to revert settings to their
default values:
Press .
When you want to cancel the setting:
Press .
3
Compensation Range Limitation Setting
Select the minimum and maximum compensa-
tion frequencies for open compensation.
Sets the minimum compensation frequen-
cy for open compensation.
Sets the maximum compensation frequen-
cy for open compensation.