5.2 Setting Basic Settings of Measurement
153
5
Chapte
r 5
ANAL
YZER Functi
on (IM3
533-01
)
Set the delay time from the initial sweep point trigger input to the start of measurement.
There are two delay settings: “Trigger Delay” and “Point Delay.”
With this item, only the setting for the trigger delay is configured.
5.2.4
Setting the Trigger Delay
When sequential sweep
When repeat sweep
When step sweep
External trigger
End of measurement
End of measurement
End of measurement
End of measurement
End of measurement
Internal trigger
Return to measurement of sweep point 1
External trigger
External trigger
External trigger
External trigger
Trigger delay
Point delay
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
last sweep point
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
last sweep point
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
last sweep point
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......