8.1 Setting Open Circuit Compensation
222
A window such as the following will be displayed if the instrument was unable to acquire normal compensation
values. If this occurs, the acquired compensation values can be enabled by touching
.
However, those values are not guaranteed.
The open circuit compensation process is quite sensitive to noise - both noise originating externally and
induced noise.Therefore, if open circuit compensation has been interrupted with a fault, you should check
the following points before starting the compensation process again:
• Check that the test cables are properly connected.
• Check that nothing is connected to the test cables. (Open circuit compensation cannot be performed
while any test sample is connected to the test cables.)
• Check that the test leads are arranged as closely as possible to their configuration in which measure-
ment will be performed.
• During the compensation process, be sure not to disturb the test cables or to move your hand near
them.
• Execute the guarding process.
"Appendix2 Measurement of High Impedance Components"(p. A3)
When Normal Compensation Values were Not Acquired
Screen when Normal Compensation Values were Not Acquired
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......