4.2 Setting Basic Settings of Measurement Conditions
53
4
Chapte
r 4
LCR Function
When the measurement value is outside the guaranteed accuracy range, the following icon
appears at the top of the screen.
In this case, you should consider the following possible causes, and you should either
change the test conditions while checking the accuracy assured ranges "14.2 Measure-
ment Range and Accuracy" (p. 342), or you should consider the measured values as values
for reference.
• Perhaps the test signal level is too low, increase the test signal level.
• If the current measurement range (during HOLD setting) is not appropriate, set again in the AUTO
range, or change the range by manual.
3
To select the measurement signal mode.
Open circuit voltage (V) mode (p. 54)
Constant voltage (CV) mode (p. 54)
Constant current (CC) mode (p. 55)
Measurement Signal Mode Setting
4
Use
or
to enter the voltage or current
value.
Measurement Signal Level Setting
See
"For setting range and accuracy" (p. 54)
Normal measurement mode
Measurement signal mode
Settable range
V, CV
0.005 V to 5.000 V
CC
0.01 mA to 50.00 mA
Low Z high accuracy mode
Measurement signal mode
Settable range
V, CV
0.005 V to 2.500 V
CC
0.01 mA to 100.00 mA
The accuracy of testing varies according to the test
signal level.
See
"14.2 Measurement Range and Accuracy" (p. 342)
5
Press
to close the setting screen.
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......