8.3 Compensating Values to Match Reference Values (Load Compensation)
240
Setting the DC Bias
• If the compensation frequency, compensation range, and compensation signal level are
not set, the DC bias setting cannot be set.
• When DC is selected for the compensation frequency setting, the DC bias setting cannot
be set.
If you make a mistake during input:
press
to cancel the input and start again.
DC Bias Setting
2.
Select ON/OFF for the DC bias.
Sets the DC bias to OFF.
Sets the DC bias to ON.
3.
Use
or
to enter the DC bias value.
Settable range: -5.00 V to 5.00 V (Normal mode)
-2.50 V to 2.50 V (low Z high accuracy
mode)
4.
Press
to close the setting screen.
Load Compensation Setting
1.
Press
.