17.6 Input Modules Specifications
309
Chapt
er
17
S
peci
fic
at
ions
17
*: During current measurement, add the accuracy and characeristic of clamp sensor using with Model 8971.
Storage temperature and humid-
ity ranges
Same as the Memory HiCorder in which the 8970 is installed
Operating environment
Same as the Memory HiCorder in which the 8970 is installed
Applicable standards
Safety EN 61010, EMC EN 61326 Class A
Dimensions and mass
Approx. 106W x 19.8H x 196.5D mm (4.17"W x 0.78"H x 7.74"D) (excluding projections),
Approx. 250 g (8.8 oz.)
17.6.6 8971 Current Unit
Temperature and humidity range
for guaranteed accuracy
23 ± 5°C (73 ± 9F°), 20 to 80% RH (when zero adjustment is executed 30 minutes after power on)
Guaranteed accuracy period
1 year
Number of input channels
2 channels
Applicable current sensors
9272-10, 9277, 9278, 9279, 9709 , CT6862, CT6863
(To connect with Model 8971, using Model 9318 Conversion Cable)
Measurement ranges
When using the Model 9272-10 (20A), 9277: 100m, 200m, 500m, 1, 2, 5 A/div
When using the Model 9272-10 (200A), 9278,CT6863: 1, 2, 5, 10, 20, 50 A/div
When using the Model 9279, 9709: 2, 5, 10, 20, 50, 100 A/div
When using the Model CT6862: 200m, 500m, 1, 2, 5, 10 A/div
Measurement accuracy*
±0.65%f.s. (filter 5 Hz ON)
±0.85%f.s. (filter 5 Hz ON, for using with Model 9278 or 9279)
RMS accuracy*
±1%f.s. (DC, 30 to 1 kHz), ±3%f.s. (1 kHz to 10 kHz), (Sine wave input, filter 5 Hz ON)
Response time*
100 ms (rise 0
→
90%f.s.)
Crest factor
2
Temperature characteristic*
±0.075%f.s./°C
Frequency characteristic*
DC Coupling: DC to 100 kHz ±3 dB
AC Coupling: 7 Hz to 100 kHz ±3 dB (low cut-off frequency 7 Hz±50%)
Noise*
10 mAp-pmax (sensitivity range, with input shorted) (for 20 A/2V range)
Low-pass filter
OFF, 5, 50, 500, 5k, 50 k±50%(Hz)-3 dB
Input type
Unbalanced (Not insulated)
Input coupling
AC/DC/GND
Input resistance
1M
Ω
±1%
A/D resolution
12 bit
Maximum sampling rate
1 MS/s
Input terminals
Sensor connector HR10A-10R-S (HIROSE)
Operating temperature and hu-
midity ranges
Same as the Memory HiCorder in which the 8971 is installed
Operating environment
Same as the Memory HiCorder in which the 8971 is installed
Storage temperature and humid-
ity ranges
-10 to 50°C (14 to 122°F), 80% RH or less (no condensation)
Dimensions
Approx. 106W x 19.8H x 196.5D mm (4.17"W x 0.78"H x 7.74"D) (excluding projections)
Mass
Approx. 250 g (8.8 oz.)
Applicable Standards
Safety EN 61010
EMC
EN 61326 Class A
Accessories
For connecting the clamp sensor: Model 9318 Conversion Cable x 2, Instruction Manual
Number of the 8971 available for
Memory HiCoder
Max. 4 unit (for Model 8847)
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......