12.2 Display Settings
204
12.2 Display Settings
Procedure
To open the screen: Press the
STATUS
key
→
[Memory Div]
sheet
1
To display any block on the Waveform screen
Set the display blocks
Set after measurement is complete.(This can also be set on the
Waveform screen (
Move the cursor to the
[Display Block]
item.
Set the number of blocks to display on the Waveform screen.
2
To display multiple blocks as overlaid waveforms
Enable the Reference Block function
Move the cursor to the
[Ref Block]
item.
Select
[On]
.
3
(When Reference Blocks are enabled [On])
Select whether to reference every block
To overlay all waveforms, select
[All Blks On]
.
To overlay selected waveforms, move the cursor to the number col-
umn of the reference block and select the block number.
Select
4
To display every block as its waveform is acquired
Enable tracking waveform display.
Move the cursor to the
[Wave Display]
item.
Select
[On]
.
Viewing Memory Division waveforms on the Waveform screen
See:
"7.8 Seeing Block Waveforms" (
Off
Reference Blocks are not displayed (default setting)
On
Reference Blocks overlay Display blocks on the display.
All Blks Off
Set reference to all blocks to Off.
All Blks On
Set reference to all blocks to On.
Ref On-Off
Set On or Off. If [On] is selected, the block frame
of the selected block number is displayed as a
green square.
↑↓
Select a block.
Off
The waveform of the specified Display Block is displayed
after recording the specified number of Used Blocks. (de-
fault setting)
On
Waveforms are displayed one block at a time as they are
acquired at each trigger event.
Display Block (Green)
Reference Block (Light Green)
Measurement data is recorded at the
colored positions.
When the Display Block is 5 and the Ref-
erence Block is 3
Reference Block No.
Enabling the Trace Waveform display
lengthens dead time.
About Dead Time:
See:
"Difference Between Dead Times
During Normal and Memory Division
Recording" (
p.206)
Even if the Roll Mode is enabled (other
than Off), it is not usable when the Trace
Waveform display is disabled.
Reference Block Selection
Reference Blocks can also be selected
and deselected in the [
Ref Block]
item on
the
[List]
display.
See:
"Getting Details on Each Block:" (
p.205)
2
4
1
Trigger time
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......