13.8 Analysis with the Waveform Screen
234
The FFT function can specify the calculation start position for waveforms measured by the memory
function before calculation.
Operation differs by calculation execution settings.
"Trigger Modes and Averaging" (
•Calculation execution:
[Single]
Analysis is performed once on the specified number of analysis points beginning with the specified
starting point, and analysis results are displayed.
This is convenient for analyzing only a specific range. However, if averaging is enabled, analysis
repeats for the specified averaging count.
•Calculation execution:
[Repeat]
Analysis is performed repeatedly on the specified number of analysis points beginning with the speci-
fied starting point and ending with end of waveform data, and final analysis results are displayed
(Calculation is performed for the number of specified points so a good end point becomes the final
analysis result.)
Verifying the analysis starting point while viewing analysis data
13.8 Analysis with the Waveform Screen
13.8.1 Analyzing after Specifying an Analysis Starting
Point
Display the waveform with the Memory func-
tion.
Set the
[Reference]
item to
[From Memory]
.
Set analysis conditions such as the analysis mode and
number of analysis points (these can also be set on the
Waveform screen).
Press the
DISP
key to display the Waveform
screen.
The one-time calculation range is displayed for the mem-
ory waveform.
1
Press the
STATUS
key to display the
[Status]
sheet of the Status
screen.
4
3
Switch to FFT function.
2
5
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......