17.2 Measurement Specifications
298
17.2 Measurement Specifications
17.2.1 Memory Function
Timebase
5, 10, 20, 50, 100, 200, 500
μ
s/div
1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5 min/div
External sampling (100 S/div, adjustable settings)
Time axis resolution
100 points/div
Sampling period
1/100th of timebase
Recording length
•
Fixed recording length
25, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000 div
50000 div (2, 4, or 8 channel mode)
100000 div (2 or 4 channel mode)
200000 div (2 channel mode)
•
Adjustable recording length
Can be set in units of divisions (up to 320,000div)
Screen and Printing Settings
1, 2, 4, 8 or X-Y (1 or 4) screens
Interpolation function
Line (exc. X-Y), dot or line (with X-Y)
Waveform scrolling
Left-right scrolling by Jog and Shuttle knobs, Backward scrolling is available during Roll Mode dis-
play
Overlay function
Auto: Always overdraws when starting, and clears by restarting
Manual: Overdraws waveforms only when necessary, and clears as needed
Automatic saving
Automatically saves data on CF card or hard disk after measuring (binary or text)
When sampling is slow, saving starts during recording.
Auto Print
Automatically prints recording waveforms after the measurement (When slow sampling is used,
printing is available even while recording.)
Manual Print
By the
key (prints up to 50 mm/s, depending on print coverage)
Selection printing
Included Prints the waveform between A/B cursors (Print All and Print Selection)
Report printing
Included
Channel modes
Selectable 2, 4, 8 or 16 channel modes
17.2.2 Recorder Function
Timebase
10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5, 10, 30 min/div, 1 h/div
Time axis resolution
100 points/div
Sampling period
1, 10, 100
μ
s
1, 10, 100 ms (not more than 1/100th of the selected timebase)
Recording length
Fixed recording length: 25, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000 div
Adjustable recording length: Can be set in units of divisions (up to 20,000div)
Continuous
Screen and Printing Set-
tings
1, 2, 4 or 8 screens
Waveform Storage
The most recent 20,000 divisions of measurement data is retained in internal memory
Waveform Scrolling
Left-right scrolling by Jog and Shuttle knobs. Including backward scrolling during measurement
Automatic saving
Automatically saves data on CF card or hard disk after measuring (binary or text)
When sampling is slow, saving starts during recording.
Realtime printing
Available with 500 ms/div and slower recording (start and stop printing by
F
key while recording),
With 10 to 200 ms/div recording, prints after recording stops (except during Continuous Record-
ing), During continuous recording at 10 to 200 ms/div, print manually after measurement stops
Manual Print
By the
key (prints up to 50 mm/s, depending on print coverage)
Selection printing
Included Prints the waveform between A/B cursors (Print All and Print Selection)
Logging recording
None
Report printing
Included
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......