8.10 Making Detailed Settings for Input Modules
148
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (
Mode
There is no need to change the setting since it is set when the clamp sensor is
automatically recognized.
Select
Range
Change the measurement of the 8971 Current Unit between DC and RMS.
Select
8.10.6 Settings for the 8971 Current Unit
Selections
Description
20A/2V
Set when a 9272-10 (20 A range) or 9277 Clamp Sensor is con-
nected. (default setting)
200A/2V
Set when a 9272-10 (200 A range), 9277, or CT6863 Clamp Sen-
sor is connected.
50A/2V
Set when a CT6862 Clamp Sensor is connected.
500A/2V
Set when a 9279 or 9209 Clamp Sensor is connected.
Selections
Description
DC
Current measurement (default setting)
RMS
RMS measurement
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......