Troubleshooting
73/90
9
Troubleshooting
General
Ø
Check whether the prerequisites for the operation of the analyzer device
NANL-B500G-RE are fulfilled:
Further information to this you find in section
Link is not recognized
Note:
For 10-Mbit/s-connections and for increased bus load the
can
not correctly detect the link. If the link is not established (Link
DOWN), for troubleshooting the speed must be set manually to 10
Mbit/s.
LINK-LED
Ø
Check using the LINK LED status if a connection to the Ethernet is
established.
Further information to this you find in chapter
Cable
Ø
Check that the pin occupation of the used cable is correct.
Analyzer device NANL-B500G-RE
Ø
Check that the power supply of the device is connected to the mains
voltage.
Host connecting interference (NANL-B500G-RE)
The Ethernet connection from the analyzer device NANL-B500G-RE device
to the PC is a saved connection without loss. Transmission errors on the
host Gigabit interface cable (such as EMC interference or bad connections)
do not cause any data loss. However, to strong interference can lead to a
disconnection.
Device is not accessible (NANL-B500G-RE)
Note:
In case of problems due to incorrect settings of the IP address (e. g.
device is not accessible), restart your device. While turning on your
device press the REC button for about 1 minute. Then the default
setting for the IP address of your device (DHCP server mode) is
restored.