Standard deviation (ppm) is used to further ensure that the measurement is stable
and within spec.
Sample Size is used to specify the number of measurements to be used for
calculating the coefficients.
Maximum Sample Count is set to specify the total number of measurements to be
taken before the process is interrupted.
It should be noted that the instrument may not reach a stable state if it is running
in an environment which has climate fluctuations or high levels of electrical or
other types of interference. This will result in the 6530 running until the
Maximum Sample Count is reached for each test that cannot satisfy the criteria set
in Standard Deviation and Variance. In some cases it may be best to change the
variance or the standard deviation from the system default values.
System Default Values:
Variance
Standard Deviation
Sample Size
200.00 (ppm)
100.00 (ppm)
30
Maximum Sample Count
1000
4.9.2.4.Sensors
This section allows setting the gain and Offset Voltage for the sensors. The gain is
specified in mV per each unit of measurement.
N.B. The default parameters are set for the Guildline environmental sensors
option.
4.9.2.4.1.Temperature