7. TROUBLE SHOOTING AND MAINTENANCE
7.1.
PREVENTATIVE MAINTENANCE
Preventative maintenance consists of cleaning and visual inspection of the instrument.
Preventative maintenance performed on a regular basis will prevent instrument breakdown and
will improve the reliability of this instrument. The severity of the environment to which the
6530 is subjected determines the frequency of maintenance. A convenient time to perform
preventative maintenance is preceding recalibration of the instrument. Do not open the
instrument for any of the described maintenance activities as removal the protective covering of
the 6530 may cause rather than prevent problems.
The 6530 should be cleaned as often as operating conditions require. Accumulation of dirt in the
instrument air filter can cause overheating and component breakdown due to improper airflow.
These filters should be vacuumed out to keep a clear airflow. The dress skins also provide
protection against dust in the interior of the instrument. Operation without these panels in place
necessitates more frequent cleaning. However, it is recommended that you remove the dress
skins to allow better access for vacuuming the air filter
Periodically inspect the instrument for general cleanliness.
CAUTION
Avoid the use of chemical cleaning agents that might damage the plastics used in this
instrument. In particular, avoid chemicals that contain benzene toluene, xylene or similar
solvents.
Periodically check the diagnostics utility in the 6530 and confirm that there are no error
conditions displayed.
Routinely run the artifact calibration procedure to ensure the 6530 is operating to full
specifications. See Section 6.
If the instrument is not correctly functioning remove the cover and clean out any accumulated
dust with a soft brush, at the same time check for discoloured or damaged wiring. Check all
screws and hardware for tightness. This should only occur as a corrective measure.
Note: A grounding wrist strap must be used to prevent electro-static discharge to sensitive
components whenever the top cover is removed.