![Guildline 6530 Series Скачать руководство пользователя страница 143](http://html1.mh-extra.com/html/guildline/6530-series/6530-series_operators-manual_2278476143.webp)
Section 8
OM6530-C1-00
19 November, 2019
8-2
8.2.
RESISTANCE MEASUREMENT SPECIFICATIONS
Note: The uncertainties listed in Table 8-2 are applicable after a four-hour warm-up period when
using the autoranging mode of operation and when the current is no less than one picoampere
through the unknown resistor. Many types of high value resistors can be difficult to measure
accurately with the 6530 in autoreverse mode because their actual resistance value changes
slowly for a period of time after a polarity reversal. They can however be measured to the full
6530 accuracy by allowing sufficient settling time between polarity reversals (soak time). This
is done under manual control or through adjusting the 6530 parameters. The accuracy is
traceable to the International System of Units (SI) through NRCC (Canada) or other National
Metrology Institutes.
24 Hour Bridge Mode ± µΩ/Ω (ppm) 23
C ± 5
C
Range (Ω)
Range (Volts)
Base
XR
XP
XPR
90 k to 200 k
1 V
NA
50
NA
40
200 k to 2 M
1 V
NA
15
NA
8
2 M to 20 M
1 V
25
15
8
8
20 M to 200 M
1 V to 10 V
25
15
8
8
200 M to 2 G
1 V to 100 V
25
15
8
8
2 G to 20 G
1 V to 100 0V
25
20
10
10
20 G to 200 G
10 V to 1000 V
25
20
15
15
200 G to 2 T
100 V to 1000 V
80
70
50
50
2 T to 20 T
1000 V
500
200
120
120
20 T to 200 T
1000 V
700
500
200
200
200 T to 2 P
1000 V
NA
1500
NA
800
2 P to 20 P
1000 V
NA
3500
NA
2000
12 Month Uncertainty Direct Measurement ± µΩ/Ω (ppm) 23
C ± 5
C
Range (Ω)
Range (Volts)
Base
XR
XP
XPR
90 k to 200 k
1 V
NA
200
NA
150
200 k to 2 M
1 V
NA
200
NA
150
2 M to 20 M
1 V
250
200
150
150
20 M to 200 M
1 V to 10 V
150
100
80
80
200 M to 2 G
1 V to 100 V
200
150
150
150
2 G to 20 G
10 V to 1000 V
600
500
400
400
20 G to 200 G
10 V to 1000 V
800
700
600
600
200 G to 2 T
100 V to 1000 V
1200
1100
1000
1000
2 T to 20 T
1000 V
3500
3000
2500
2500
20 T to 200 T
1000 V
6000
5000
4000
4000
200 T to 2 P
1000 V
NA
20,000
NA
15,000
2 P to 20 P
1000 V
NA
250,000
NA
200,000
Table 8-2: Resistance Measurement Uncertainty