Section 4
OM6530-C1-00
19 November, 2019
4-9
4.4.2.
Ohms Parameters Menu
The parameters menu allows the manual selection of settings.
The Max Volts limits all measurements. Max Volts is set to 30 V as power up default.
This also applies to auto ranging. This may be important to note as some resistors may be
damaged if high voltage is applied. The valid ranges are:
1 V
10 V 100 V 1000V
3 V
30 V 300 V
Selection of the Capacitor, Threshold or Test Volts command will automatically set the
measurement mode to manual and stop the measurement if it is active.
The Test Volts command selects the voltage to be used during the measurement. The
valid ranges are the same as Max Volts however it will not allow the selection to exceed
Max Volts. The Max Volts will have to be increased if the desired voltage is greater than
Max Volts. This is a deliberate limit to ensure that the voltage is not accidentally changed
above the tolerance of the resistor. In Auto Range mode the 6530 will not provide a
calibrated measurement if Max Volts is set too low.
The capacitor command selects the integrating capacitor. The valid capacitor values are:
27 pf 270 pf 2700 pf
The 27 pf and 270 pf selections are only available if the 0.1 V threshold is selected.
The Threshold command selects the threshold voltage. The valid voltages are:
0.1
V 1.0 V 10.0 V