Section 5
OM6530-C1-00
19 November, 2019
5-28
5.6.56.
CALibration:SYSTem:COEFficient?
Typical response:
90k to 200k, 1V, 273,273
5.6.57.
CALibration:SYSTem:PARameters
<nominal value>, <volts>, <sample count>, <sample size>, <sample average>,<sample
window>
5.6.58.
CALibration:SYSTem:PARameters?
Typical response:
90k to 200k, 1V, 90, 100, 1, 2
5.6.59.
CALibration:SYSTem:REV
<user rev>
5.6.60.
CALibration:SYSTem:REV?
Typical response:
<user rev> / <system rev>
5.6.61.
CALibration:THReshold:VOLTage
0.1
| 1.0 <calibration coefficient (ppm)>
5.6.62.
CALibration:THReshold:VOLTage?
Typical response:
0.1V, 37, 1.0V, -160
5.6.63.
CONFigure:TEST:VOLTage
START | CONTinue | DISable
When the system is in remote or lockout then it is necessary to send a
<CONTinue>
command within twenty (20) seconds when measuring resistance or resistivity. This is a
safety precaution to prevent the high voltage section from remaining active when the
remote controlling device is not in a known operating state.