Section 6
OM6530-C1-00
19 November, 2019
6-1
6. VERIFICATION AND CALIBRATION
6.1.
CALIBRATION DESCRIPTION
The procedure outlined in this section may be used to perform system verification of the 6530
TeraOhm Bridge-Meter for proper operation. This verification procedure may be performed
when the instrument is first received to ensure that no damage or maladjustment has occurred
during shipment.
6.1.1.
ENVIRONMENTAL CONDITIONS
Verification checks should be made only when the instrument is being operated within
the operating limits of temperature and humidity specified in Section 8 of this manual.
6.1.2.
INITIAL CONDITIONS
A warm-up time of at least four hours must be allowed before beginning the verification
process. If the instrument has been subjected to extremes of temperature outside the
operating limits, additional time should be allowed for the instrument components to
stabilize to their normal operating temperatures. Typically, it takes one additional hour to
stabilize a unit that has been exposed to a temperature 10
C outside the specified
temperature range.
6.1.3.
RECOMMENDED TEST EQUIPMENT
Table 6-1 lists all test equipment required for the verification of the 6530 TeraOhm
Bridge-Meter. Alternate equipment may be used as long as the substitute equipment has
specifications as good as or better than the equipment listed.
Description
Minimum Use Specifications
Recommended
Equipment
Output Voltage
Range (0 to ± 1000) DC V,
accuracy ± 0.1% of output
DC V Meter accuracy
< 60 ppm or better,
minimum 10 MΩ input
impedance
Table 6-1: Recommended Verification Test Equipment
6.1.4.
CHECK CALIBRATION REPORT
The following paragraph details the procedures to be used to check the stored calibration
coefficients against the coefficients listed on the instrument calibration report. If the
instrument has been re-calibrated or adjustments made to the instrument after the date
printed on the calibration report, the stored coefficients may not match the coefficients
listed in the report. The user should verify the instrument against the most recent
calibration report.