Section 5
OM6530-C1-00
19 November, 2019
5-17
5.6.
REMOTE COMMANDS
This section details all the valid commands that may be sent over either the IEEE-488 or the
RS-232C interface port. The responses listed below are the verbose response.
If the numeric parameter to the command is missing or unrecognizable the CME (CoMmand
Error) bit in the Event Status Register (see Figure 5-1) will be set. If the unrecognizable
command was sent over the RS-232C interface, the 6530 will respond with “Unrecognized
Command”.
If the numeric value is out of range, then the EXE (EXecution Error) bit will be set for a program
data element out of range error. If the out of range value was sent over the RS-232C interface,
the 6530 will respond with “Invalid Parameter”.
Unless otherwise indicated, the terse response is that portion of the response printed in bold.
5.6.1.
*CLS - CLEAR STATUS COMMAND
This command clears all Event Status Registers summarized in the status byte register.
All queues, except the Output Queue, that are summarized in the status byte register are
emptied. The 6530 is forced into the Operation Complete Idle State and the Operation
Complete Query Idle state.
5.6.2.
*ESE <u> - SET EVENT STATUS ENABLE REGISTER
This command sets the standard event status enable register bits. When the bits in the
Event Status Enable (ESE) register are "ANDed" with the bits in the Event Status
Register (ESR) if the result is non-zero then the Event Status Bit (ESB) in the Status Byte
(STB) register is set.
The values accepted for the *ESE command are between 0 and 255, all other values are
considered to be an error. The default value for the Event Status Enable (ESE) register at
power on is zero (0).
5.6.3.
*ESE? - EVENT STATUS ENABLE QUERY
This command reports the current value of the Event Status Enable Register. The value
returned will be between 0 and 255.
5.6.4.
*ESR? - EVENT STATUS REGISTER QUERY
This query allows the programmer to determine the current contents of the event status
register. Reading the Event Status Register clears it.