Section 4
OM6530-C1-00
19 November, 2019
4-38
Coefficient into display | Nominal Voltage |
Due to the large number of output test voltage points to be measured
during calibration, it is recommended that the numbers are entered using
one of the interface busses (GPIB or RS-232C).
4.9.2.2.4.Cal Spec.
This displays the statistics of the last time that the Utilities were used to establish
the Capacitor and Threshold coefficients. It displays the total count of samples
required to meet the specification as well as the actual variance and standard
deviation for the specified sample size.
4.9.2.2.5.System Parameters
Pressing the
<Sys Params>
key will display the system parameters stored for the
6530.
<Function key 3>
selects the resistance range.
<Function key 4>
selects
the available voltage within the resistance range. By pressing the
<Edit>
key the
numbers may be altered by pressing a new sequence of numbers followed by
pressing the
<Ok>
key. Pressing the
<PREVIOUS>
key exits to the SOFCAL
Calibrate Menu.
NOTE: Do not alter the system parameters as they will affect the accuracy of the
6530. These are factory set parameters. Access to the parameters is only
recommended in the event of a non-volatile memory failure, to confirm and, if
required, edit them to return them to the values set by the factory. These values
are supplied in hard copy on the instrument report.