Section 4
OM6530-C1-00
19 November, 2019
4-8
The setup menu allows the user to select the polarity using the polarity command
function.
Changing polarity will not stop the measurement. It is recommended that attention is paid
to the parameters for auto reverse as this will drastically affect the accuracy of the
measurement. Voltage reversals for measurements greater than 10 GΩ require longer
settling time. If the system is configured in 6530 mode and autoranging is used then the
system will automatically select appropriate parameters for sample size and sample count
based on the resistance. (see section 4.9.2.2.5 System parameter) In manual mode Auto
Reverse Sample Count and Stabilize Size (4.4.6. Timers) are used to establish these
values.