4.9.2.1.4.Zero Calibration
This is the value of the internal resistance presented by the 6530 between the
input connector and ground. The 6530 maintains a stored value for this resistance
in its memory and subtracts it from all resistance measurements before the final
value of the unknown resistance is put on the front panel display or made
available at either control bus. If the ZERO CAL resistance should change or if
the stored value should become corrupted in the memory, the 6530 is capable of
re-measuring the true value and entering the true value into memory.
<Calc>
starts the calibration procedure.
If the specified variance or standard deviation is exceeded then the sample count
is reset to zero and the failed spec is displayed on the 5
th
l,ine of the display. Line
6 of the display consists of the total measurement count, the number of samples
which have currently been taken for the valid sample set and the current variance
(ppm) and standard deviation (ppm) for the valid sample set.
When the number of samples specified has been measured and have not exceeded
the specification then the calibration routine has successfully completed and the