2. INSTALLATION
2.1.
INSTALLATION
This instrument was thoroughly tested and inspected before shipment and should be free from
damage when received. Inspect it carefully, verify that all items on the packing list are present
and check the instrument operation as soon as possible. Refer to the warranty card at the front of
this manual if any damage or deficiencies are found.
The 6530 TeraOhm Bridge-Meter is an instrument intended to be used in a laboratory
environment and is specified to be operated within an environmental temperature range of 23
C
±5
C with humidity levels in the range 40 % rH ± 10 % rH. Optimum performance is achieved
when the environmental temperature range is (23 ± 2)
C. Higher humidity can degrade the
accuracy of the instrument. The 6530 must be mounted with an angle of inclination of no more
than 30
. Where the TeraOhm Bridge-Meter is to be used in a rack, attach the mounting brackets
provided. To attach the rack mounting flanges (brackets), the original screws holding the
handles to the instrument are removed and the flanges attached over the handles with the longer
screws supplied. The instrument has to be supported in the rack/cabinet with adjustable support
angles or a support bar. In case of interference with other equipment mounted directly below the
instrument, the 4 feet must be removed. Install the unit in the rack.
2.2.
PRELIMINARIES
The model 6530 has been shipped with the line input voltage set to 240 volts with the fuse
removed. The line input selectors must be set to the correct line voltage before power is applied
to the instrument. Remove the warning label positioned across the power entry only after setting
the proper operating voltage. The line input voltage selection must be set correctly. The settings
available are 100 V, 120 V, 220 V, and 240 V. Figure 2-1 details the line input voltage selector.
The instrument is supplied with a North American style line cord, unless otherwise specified at
time of order. Ensure that the line cord is plugged into a wall socket or extension cord that has a
protective or safety ground. Where 3-contact power supply outputs are not available, a suitable
protective ground connection must be made before switching the instrument power on. Any
interruption of the protective ground may possibly render the instrument unsafe.
To set the correct line input voltage pry open the power receptacle on the rear panel as shown in
Figure 2-1.