Section 4
OM6530-C1-00
19 November, 2019
4-14
<Pressure>
selects and removes the atmospheric pressure data from the trace.
<State>
selects and removes the machine state data from the trace.
4.4.6.
Timers
Delay is the time to wait between each sample. It is an internal time based sample trigger.
Soak Time is the initial settling wait time after a change in voltage or polarity.
Sensor Time is the wait time between sensor measurements.
4.4.7.
Profiles
Profiles allow the user to select standard default setups or to create individual user
profiles. 36 profiles are available in the 6530; 3 are fixed (LOCAL, GPIB,
RS232) and 33 are user definable. The remaining 33 profiles are stored in non-
volatile memory and will be retain by the 6530 even after powering off.
<Select>
allows the selection and deletion of profiles.
Содержание 6530 Series
Страница 2: ......
Страница 12: ...OM6530 C1 00 19 November 2019 x...
Страница 18: ...Section 1 OM6530 C1 00 19 November 2019 1 6 Figure 1 4 6530 Operational Integrator...