Section 4
OM6530-C1-00
19 November, 2019
4-16
4.4.8.
View
The
View Menu
allows the user to view the trace data in graphical and text
format. The last 1000 entries are kept in memory and the individual entries can be
viewed. A summarization of the data is kept from the last
<Clear Sum>
point. A
graphical summary is also maintained by averaging all samples within a 180 point
array.
4.4.8.1.Summary
The
Count
entry indicates how many actual values have been averaged within the
summary graph.
<Refresh>
updates the detail and summary values.
<Detail>
switches to the Detail display.
<Clear Det>
clears the trace buffer. It requires a confirmation to prevent
accidental clearing.
<Clear Sum>
clears the summary buffer