75
Appendix 4: Connecting of Pt1000 RTD probe
It is recommended to use the intelligent probes Pt1000/M with
the Multilogger. This product line includes the memory with probe
information and data for automatic correction of the cable impedance.
The
Pt1000/M probes are interchangeable without need to consider the cable
length. May be that the connection of the standardize RTD Pt1000 probe
will be necessary. In this case you can connect this probe in
accordance with drawing shown below. Use four-pin MiniDIN connector
(
Guilcor
order code
AR006465
). The input of the Multilogger have to
be
set to
“
Custom Pt1000
”
using the SW (instead of Pt1000/M). For
the correction of cable impedance the
User calibration
can be used.
Connecting a Pt1000 temperature
probe to the MiniDin connector:
Содержание AR004926
Страница 5: ...5 Multilogger drawing only certain models...
Страница 10: ...10 The way of mounting the holder to Multilogger and mounting holes dimensions MLP001 AR006460...
Страница 11: ...11 AR006462...
Страница 19: ...19 Connection of the transmitter with the two current outputs passive...
Страница 21: ...21...
Страница 73: ...73 Appendix 2 Output and power supply circuits...
Страница 74: ...74 Appendix 3 Input circuits...