63
The following devices can be connected to this input:
Thermocouple of the “K” type
(NiCr-Ni)
Measured quantity:
temperature
Range:
(-
200 to 1,300) °C
Accuracy (without probes)
± (0.3% from the measured value + 1.5 °C)
Resolution:
better than 0.1 °C *)
Cold junction:
compensated in the temperature range of (-10
to 60) °C
______________________________________________________________________________________
*)
to achieve this value, the corresponding number of
decimals must be set
Thermocouple of the “J” type
(Fe-Co)
Measured quantity:
temperature
Range:
(-
200 to 750) °C
Accuracy (without probes)
± (0.3% from the measured value + 1.5 °C)
Resolution:
better
than 0.1 °C *)
Cold junction:
compensated in the temperature range of (-10
to 60) °C
______________________________________________________________________________________
*)
to achieve this value, the corresponding number of
decimals must be set
Thermocouple of the “S” type
(Pt10%Rh-Pt)
Measured quantity:
temperature
Range:
(0 to 1,700) °C
Accuracy (without probes)
± (0.3% from the measured value + 1.5 °C)